Additional Digitize Default Settings
On entering digitize mode, the DMM configures these settings:
l If trigger source is set to manual it is changed to auto. (external and level remain as is.)
l Limits mode is turned off.
l Scaling is turned off.
l Statistic and histogram are put in post-processing mode (computed after digitize is complete).
l On the selected function (DCV or DCI) and for the new function if changed:
l Autorange is turned off.
l Autozero is turned off.
l NPLC and aperture are set to their minimum values.
l If trigger source is external or level, the pretrigger count is set to the digitize pretrigger count setting
(default of 0).
l Samples per trigger is set according to the digitize duration (time or samples).
l Sample timer is put in timer (not immediate) mode and Sample timer is set according to the digitize
sample rate or sample interval.
l Trend Chart mode is changed to bucketized when data logging to a file.
l On return to Continuous mode, settings are left as done in data log or digitize mode except:
l Sample source is set to immediate.
l Pretrigger count is set to 0.
l Samples per trigger is set to 1.
l Trigger count is set to infinite.
Keysight Truevolt Series Operating and Service Guide