Front Panel Menu Reference
The following table summarizes the front panel keys and the menu structure.
Key Purpose
Configure DC voltage measurements, including DCV ratio measurements:
Range: Autorange (default), 100 mV, 1 V, 10 V, 100 V, or 750 V
Aperture NPLC: 0.02, 0.2, 1, 10, 100. Default: 10 (34460A/61A)
0.02, 0.06, 0.2, 1, 10, 100. Default: 10 (34465A/70A without DIG option)
0.001, 0.002, 0.006, 0.02, 0.06, .2, 1, 10, 100. Default: 10 (34465A/70A with
DIG option)
See Range, Resolution and NPLC for more information.
Aperture Time (applies only to the 34465A and 34470A): (Without the
DIGoption) 200µs to 1 s (2 µs resolution), default: 100 ms. (With the DIG option)
20 µs to 1 s (2 µs resolution), default: 100 ms.
Auto Zero: Off or On (default)
Input Z: 10 (default) or HighZ (> 10 GΩ)
DCV Ratio: Off (default) or On
Configure DC current measurements:
Terminals: 3 A or 10 A
Range: Auto, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, or 10 A (when terminals set
to 10 A). The 34465A and 34470A have additional 1 µA and 10 µA DC current
ranges.
Aperture NPLC: 0.02, 0.2, 1, 10, 100. Default: 10 (34460A/61A)
0.02, 0.06, 0.2, 1, 10, 100. Default: 10 (34465A/70A without DIG option)
0.001, 0.002, 0.006, 0.02, 0.06, .2, 1, 10, 100. Default: 10 (34465A/70A with
DIG option)
See Range, Resolution and NPLC for more information.
Aperture Time (applies only to the 34465A and 34470A): (Without the
DIGoption) 200µs to 1 s (2 µs resolution), default: 100 ms. (With the DIG option)
20 µs to 1 s (2 µs resolution), default: 100 ms.
Auto Zero: Off or On (default)
Configure AC voltage measurements:
Range: Autorange (default), 100 mV, 1 V, 10 V, 100 V, or 750 V
Filter: >3 Hz, >20 Hz, >200 Hz
Configure AC current measurements:
Terminals: 3 A or 10 A
Range: Auto, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A, or 10 A (when terminals set
to 10 A)
Filter: >3 Hz, >20 Hz, >200 Hz