EasyManua.ls Logo

Keysight 34461A - System Subsystem LOCK Commands

Keysight 34461A
583 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SYSTem Subsystem LOCK Commands
This subsystem locks and releases instrument I/O interfaces, as shown in the following example:
Initial State = unlocked, Count = 0
(FROM USB) SYST:LOCK:REQ? returns "1" (request successful)
State = locked, Count = 1
(FROM GPIB) SYST:LOCK:REQ? returns "0" because USB has lock
State = locked, Count = 1
(FROM USB) SYST:LOCK:REQ? returns "1" (request successful)
State = locked, Count = 2
(FROM USB) SYST:LOCK:REL
State = locked, Count = 1
(FROM USB) SYST:LOCK:REL
State = unlocked, Count = 0
For each successful lock request, a lock release is required. Two requests require two releases.
Command Summary
SYSTem:LOCK:NAME?
SYSTem:LOCK:OWNer?
SYSTem:LOCK:RELease
SYSTem:LOCK:REQuest?