Power-On and Reset State
These tables show factory default settings. Parameters marked with a bullet (•) are nonvolatile and are not
affected by power-on or system reset. Other parameters are volatile and are reset to the indicated values
at power-on or after *RST or SYSTem:PRESet.
The power-on/reset state may differ from these values if you have enabled
power-on state recall mode.
Factory Default Settings
Measurement Configuration Factory Setting
Function DC Volts
Range Autorange (for all functions)
Integration Time NPLC On, 10 power line cycles (PLC)*
Resolution Resolution equivalent to 10 PLC
Autozero On*
Aperture (34465A/70A only)* Off, 100 ms
Frequency or Period Aperture (Gate Time) 100 ms*
Input Impedance 10 (fixed for all DC voltage ranges)
AC Input Filter (bandwidth) 20 Hz (medium filter)
Temperature Transducer Type FRTD
RTD Reference Resistance 100
Nulls (individual functions) Off; reference = 0.0, auto reference enabled
(for all measurement functions)
Current Terminals 3 A
Secondary Measurements Off
Screen Shot Format PNG
Calibration Value 0.0
Current Switch Mode (34465A/70A only) Continuous
Resistance (2- and 4-wire) (34465A/70A only):
Offset Compensation Off
Low Power Measurements Off
Temperature (34465A/70A only):
Offset Compensation (2- and 4- wire RTD) Off
Low Power Measurements(2- and 4- wire RTD or thermistor) Off
Thermocouple (34465A/70A only):
Check Off
Reference Junction Temperature, Type, Adjust C, Internal, C
Thermocouple Type J
* For all DC measurement functions.
SCPI Programming Reference
474
Keysight Truevolt Series Operating and Service Guide