Automatic Trigger Delays
If TRIGger:DELay:AUTO is on, the instrument selects the trigger delay for you, as shown in the tables
below.
All frequency and period measurements have an automatic trigger delay of 1 second.
The continuity and diode functions ignore the trigger delay.
The 2-wire and 4-wire temperature measurements use the corresponding 2-wire and 4-wire
resistance measurement delays.
DC Voltage Default Delays
PLC
0.001
1
0.002
1
0.006
1
0.02
0.06
2
0.2 1 10 100
Aperture (60 Hz Power)
20 µs
1
40 µs
1
100 µs
1
0.3 ms
1 ms
2
3 ms
16.7 ms 0.167 s 1.67 s
Aperture (50 Hz Power) 20 ms 0.2 s 2 s
Range Delay
All 100 µs 130 µs 160 µs
1
34465A/70A with DIG option only.
2
34465A/70A only.
DC Current Default Delays
PLC
0.001
1
0.002
1
0.006
1
0.02
0.06
2
0.2 1 10 100
Aperture (60 Hz Power)
20 µs
1
40 µs
1
100 µs
1
0.3 ms
1 ms
2
3 ms
16.7 ms 0.167 s 1.67 s
Aperture (50 Hz Power) 20 ms 0.2 s 2 s
Range Delay
All 1 ms 1.5 ms
1
34465A/70A with DIG option only.
2
34465A/70A only.
SCPI Programming Reference
456
Keysight Truevolt Series Operating and Service Guide