SYSTem:ACALibration:TIME?
Returns the time of the last autocalibration in the format hh,mm,ss.sss.
Parameter Typical Return
(none) 20,15,30.000
Return the calibration time:
SYST:ACAL:TIME?
l (34465A/70A only)
l The date is based on the instrument's real-time clock. Set the date for the instrument's real-time clock
with SYSTem:DATE.
l The real-time clock does not adjust itself for time zone changes or daylight savings time. Be sure to set
the date and time when you first receive your instrument.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).
l Use *CAL? to perform autocal.
See Also
SYSTem:TIME
SYSTem:BEEPer[:IMMediate]
Issues a single beep.
Parameter Typical Return
(none) (none)
Issue a single beep:
SYST:BEEP
l Programmed beeps may be useful for program development and troubleshooting.
l You can issue a beep with this command regardless of the current beeper state (SYSTem:BEEPer-
:STATe).