PERFORMANCE TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
Date _________________________
DC VOLTAGE TESTS
CAL? 59 _________________________
TEMP? _________________________
Difference _________________________ –(must be less than 5 degrees C)
Perform an ACAL DCV
Test #
3458A 
input
3458A 
range
Transfer 
standard 
read ing
Unit 
under test 
reading
Difference
Limit 
(Std)
Limit 
(Opt 002)
Pass Fail
Offset test (NOTE: Math null is disabled)
1 Short 100 mV N/A __________ N/A 000.00106 mV 000.00106 mV __________ __________
2 Short 1 V N/A __________ N/A 0.00000106 V 0.00000106 V __________ __________
3 Short 10 V N/A __________ N/A 00.0000023 V 00.0000023 V __________ __________
4 Short 100 V N/A __________ N/A 000.000036 V 000.000036 V __________ __________
5 Short 1000 V N/A __________ N/A 0000.00010 V 0000.00010 V __________ __________
Gain test
1 100 mV 100 mV __________ __________ __________ 000.00212 mV 000.00188 mV __________ __________
2 1 V 1 V __________ __________ __________ 0.00000998 V 0.00000740 V __________ __________
3 1 V 10 V __________ __________ __________ 00.0000111 V 00.0000085 V __________ __________
4 –1 V 10 V __________ __________ __________ 00.0000111 V 00.0000085 V __________ __________
5 –10 V 10 V __________ __________ __________ 00.0000892 V 00.0000624 V __________ __________
6 10 V 10 V __________ __________ __________ 00.0000892 V 00.0000624 V __________ __________
7 100 V 100 V __________ __________ __________ 000.001114 V 000.000853 V __________ __________
8 1000 V 1000 V __________ __________ __________ 0000.02396 V 0000.01934 V __________ __________