PERFORMANCE TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Multimeter
Serial number _________________________ Test performed by _________________________
Date _________________________
DC CURRENT TESTS
CAL? 59 _________________________
TEMP? _________________________
Difference _________________________ –(must be less than 5 degrees C)
Perform an ACAL OHMS
Test #
3458A 
input
3458A 
range
Transfer 
standard 
read ing
Unit 
under test 
reading
Difference
Limit 
(Std)
Limit 
(Opt 002)
Pass Fail
Offset test (NOTE: Math null is disabled)
1 Open 100 µA N/A __________ N/A 000.00095 µA 000.00095 µA __________ __________
2 Open 1 mA N/A __________ N/A 0.0000065 mA 0.0000065 mA __________ __________
3 Open 10 mA N/A __________ N/A 00.000065 mA 00.000065 mA __________ __________
4 Open 100 mA N/A __________ N/A 000.00065 mA 000.00065 mA __________ __________
5 Open 1 A N/A __________ N/A 0.0000115 A 0.0000115 A __________ __________
Gain test
1 100 µA 100 µA __________ __________ __________ 000.00356 µA 000.00356 µA __________ __________
2 1 mA 1 mA __________ __________ __________ 0.0000323 mA 0.0000323 mA __________ __________
3 10 mA 10 mA __________ __________ __________ 00.000323 mA 00.000323 mA __________ __________
4 100 mA 100 mA __________ __________ __________ 000.00489 mA 000.00489 mA __________ __________
5 1 A 1 A __________ __________ __________ 0.0001349 A 0.0001349 A __________ __________