EasyManua.ls Logo

Keysight 34980A - Environmental Conditions

Keysight 34980A
67 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Keysight 34980A Getting Started Guide
Environmental Conditions
Keysight 34980A is designed for indoor use in an installation category II and low
condensation environment. Table below shows the general environmental
conditions for this instrument. Refer to the product data sheet at
https://literature.cdn.keysight.com/litweb/pdf/5989-1437EN.pdf for more
information on the instrument general specifications.
General specifications Requirement
Temperature
Operating condition: 0°C to 55°C
Storage condition: -40°C to 70°C
Humidity
Maximum Relative Humidity (non-condensing): 80% RH
up to 40°C, decreases linearly to 37% RH at 55°C
Altitude Up to 2,000 m
Pollution degree 1 or 2
NOTE
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence (on insulation) (IEC 61010-1).
NOTE
Pollution Degree 2: Normally only non-conductive pollution occurs.
Occasionally, a temporary conductivity (leakage current between isolated
conductors) caused by condensation can be expected (IEC 61010-1).

Other manuals for Keysight 34980A

Related product manuals