132 Keysight E4980A/E4980AL Precision LCR Meter
Error Messages
Error Messages
-
-440 Query UNTERMINATED error after indefinite response
A query that requests arbitrary data response (*IDN? and *OPT? queries) is
sent before normal queries in a program message (for example, FREQ?;*IDN?
is expected, but *IDN?;FREQ? is received).
R
16 Reference Measurement aborted
Reference data measurement is aborted.
S
71 Scanner I/F not installed
Operation that requires option 301 is performed in a model that does not have
a scanner interface board (option 301).
82 Store failed
User attempts to save data into number 10 to 19 when a USB memory device
is not inserted. Prepare the USB memory device or save data into number 0 to
9.
-150 String data error
String data is improper.
-158 String data not allowed
String data is not allowed.
-138 Suffix not allowed
A suffix is not allowed for this operation.
-102 Syntax error
Unrecognized command or data type was received.
T
-223 Too many digits
The received numeric value is too long (exceeds 255 digits).
-223 Too much data
The block-, expression-, or character-string-type program data that are
received conform with the standard but exceed the amount that can be
processed under the condition of the memory or conditions specific to
memory-related devices. In this instrument, this error occurs when the number
of characters exceeds 254 in a character-string parameter.
-211 Trigger ignored