Testing Performance 2
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes Service Guide  23
 
The difference in average voltage readings should be within the test limits of 
Table 5.
If a result is not within the test limits, go to the “Troubleshooting” chapter. Then 
return here.
11 Disconnect the oscilloscope calibrator from the oscilloscope.
12 Repeat this procedure to check the DC vertical gain accuracy with the 
remaining Volts/div setting values in Table 5.
13 Finally, repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 0.5 mV/div, 1 mV/div, 2 mV/div, and 
5 mV/div, noise may be a factor. To eliminate the noise, add a BNC Tee, blocking 
capacitor, and shorting cap at the oscilloscope channel input to shunt the noise to 
ground. See Figure 2. If a BNC capacitor is not available, use an SMA blocking 
capacitor, adapter, and cap. See “Blocking capacitor and shorting cap in the 
equipment list on page 17 for details.
Figure 2 Using a Blocking Capacitor to Reduce Noise
To oscilloscope input
BNC 
shorting cap
Blocking 
Capacitor