Testing Performance 2
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes Service Guide 37
Internal Trigger Sensitivity
Generator Setting Test Limits, greater of Channel 1 Channel 2 Channel 3 Channel 4
200 MHz models: 10 MHz 0.6 div or 2.5 mV ________ ________ ________ ________
Max BW (200 MHz) 1.2 div or 5 mV ________ ________ ________ ________
100 MHz models: 10 MHz 0.6 div or 2.5 mV ________ ________ ________ ________
Max BW (100 MHz) 1.2 div or 5 mV ________ ________ ________ ________
70 MHz models: 10 MHz 0.6 div or 2.5 mV ________ ________ ________ ________
Max BW (70 MHz) 0.9 div or 3.8 mV ________ ________ ________ ________
50 MHz models: 10 MHz 0.6 div or 2.5 mV ________ ________ ________ ________
Max BW (50 MHz) 0.9 div or 3.8 mV ________ ________ ________ ________