17
Conducted Emissions Measurement Examples
Making Conducted Emission Measurements
Measuring the Quasi-peak and average values of the signals
Measurement tip
If the signals you are looking at are in the lower frequency range of the conducted 
band, 2 MHz or lower, you can reduce the stop frequency to get a closer look. Note 
that there are fewer points to view. You can add more data points using the scan 
table. The default setting in the scan table is two data points per BW or 4.5 kHz per 
point in this case since the resolution bandwidth is 9 kHz. To get more data points, 
change the points per bandwidth to four points.
Step Action Notes
1 Perform a Re-measure on 
all signals in the list
a. Press Meas Setup, Scan. 
Sequence, (Re)measure, 
(Re)measure, All Signals, 
Sweep/Control, Start.
2 Review the measurement results The delta to Limit Line values 
should all be negative. If some 
of the measurements are 
positive, there is a problem 
with conducted emissions from 
the DUT.