EasyManua.ls Logo

Keysight EXA N9010A - Page 40

Keysight EXA N9010A
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
Disturbance Analyzer (Click) Measurements
Setup Table Menu Features
Factor f
For certain types of products that must be cycled to emit discontinuous disturbances
(rather than run continuously), CISPR 14-1 requires users to operate the product over
enough cycles to product 40 clicks.
Factor f is used to calculate the click rate for these types of devices. See CISPR 14-1,
Annex A Table A.2 for the factor to use for your specific DUT.
NOTE This information is given as an example. CISPR 14-1 is the reference document for
disturbance measurement requirements. Refer to CISPR 14-1 to identify the test
requirements for your specific DUT.

Other manuals for Keysight EXA N9010A

Related product manuals