52 S-Series Oscilloscopes Service Guide
4 Troubleshooting
 
Troubleshooting Overview
The service strategy for troubleshooting Infiniium S-Series oscilloscopes is to 
isolate problems to a faulty assembly, then use the disassembly and assembly 
procedures in Chapter 5, “Replacing Assemblies”, to replace the defective 
assembly.
Read the Safety Notices in the last chapter before servicing the oscilloscope. 
Before performing any procedure, review it for any cautions and warnings.
The only equipment you need for troubleshooting to the assembly level is basic 
electronic troubleshooting tools such as a digital multimeter. If you need to 
remove and replace any assemblies, refer to Chapter 5, “Replacing Assemblies”.
A default setup is provided to return the oscilloscope to a known state. You can 
use the default setup to undo previous setups so they do not interfere with the 
current measurement. Use the default setup when a procedure requires it by 
pressing [Default Setup] on the front panel.
INJURY CAN RESULT! Use caution when the oscilloscope fan blades are exposed as 
they can cause injury.
AVOID ESD DAMAGE TO COMPONENTS! Electrostatic discharge (ESD) can damage electronic 
components. Use proper ESD precautions when doing any of the procedures in this chapter. Failure to 
follow proper ESD procedures may cause immediate failure or latent damage. Latent damage may result 
in equipment failure after a period of time. As a minimum, place the oscilloscope on a properly 
grounded ESD mat and wear a properly grounded ESD strap.
SHOCK HAZARD! Maintenance should be performed by trained service personnel. 
Lack of training and awareness could result in electrical shock or other injury. When 
maintenance can be performed without power applied, the power cord should be 
removed from the oscilloscope.