66 S-Series Oscilloscopes Service Guide
4 Troubleshooting
 
Setting up the BIOS
If the BIOS settings become corrupt, the Infiniium oscilloscope PC motherboard 
will not recognize the hard drive and the oscilloscope may not start. To configure 
the motherboard BIOS parameters to the default settings, follow these steps:
1 Connect the power cable to the oscilloscope.
2 Connect the external keyboard to the rear panel.
3 Press [Del] as soon as the following prompt appears on the bottom of the 
screen:
Press DEL to enter SETUP, F12 to select boot device.
4 Go to F3 Optimized Defaults and press [Enter]. Select Yes to load the defaults, then 
press [Enter].
5 Go to the Save and Exit Setup option and press [Enter] to save and exit the setup. 
Select Yes to save and exit, then press [Enter].