Calibration
42 Keysight M9181A Calibration and Verification Procedure
AC Current Gain and Offset Calibration Procedure
Review “Test Considerations” on page 16 and “Gain Adjustments” on page 36
before beginning this procedure.
1 Configure each function and range shown in the table below.
2 Apply the input signal shown in the “Input” column of the table. Enter the
actual applied rms input voltage (see Adjustments” on page 33).
SFP Cal Step Step Name Calibrator Input Amplitude, Function
26 DCI Offset Open
27 (Skip for M9181A - Applies only to M9182A/83A)
28 (Skip for M9181A - Applies only to M9182A/83A)
29 (Skip for M9181A - Applies only to M9182A/83A)
30 (Skip for M9181A - Applies only to M9182A/83A)
31 DCI Gain .002 A 0.002 IDC
32 DCI Gain .02 A 0.02 IDC
33 DCI Gain .2 A 0.2 IDC
34 DCI Gain 2 A 2 IDC
NOTE
The Fluke 5720A calibrator does not have the capability to source the currents
for calibration steps 27, 28, and 29. Source these currents by performing the
following steps. All other calibration steps are sourced using the Fluke 5720A
and its DCI output with the current shown in the Input section above.
NOTE
Before the ACI Offset on a given range can be performed, the ACI gain on that
range must be performed (for example, you must perform step 35 before step
39).