Verification
Keysight M9181A Calibration and Verification Procedure 13
Verification
This manual contains procedures for calibrating and verifying the performance of
the Keysight M9181A PXI 6½ Digit Multimeter (DMM).
The following table lists the test equipment recommended for calibration and
performanceverification. You may substitute calibration standards of equivalent
accuracy.
Application Recommended Equipment Accuracy Requirements
DC Voltage Fluke 5720A <1/5 instrument 1 year spec
DC Current Fluke 5720A <1/5 instrument 1 year spec
Resistance Fluke 5720A <1/5 instrument 1 year spec
AC Voltage Fluke 5720A <1/5 instrument 1 year spec
AC Current Fluke 5720A <1/5 instrument 1 year spec