Test 4: 2-Wire Ohms Function
24 Keysight M9181A Calibration and Verification Procedure
Calibrator Input
Resistance (W)
DMM Function DMM Range (W) Quick Check
1 Year Error from
Nominal (± W)
190 2W W 200 3.070E-02
1,900 2W W 2,000 2.680E-01
19,000 2W W 20,000 Q 2.680E+00
190,000 2W W 200,000 4.400E+01
1,900,000 2W W 2,000,000 1.220E+03
19,000,000 2W W 20,000,000 3.860E+04