Test 6: AC Volts Verification Test
26 Keysight M9181A Calibration and Verification Procedure
Calib. Source
Input Voltage
(VAC)
Calib. Source
Freq (Hz)
DMM
Function
DMM
Voltage Range
Quick Check
1 Year Error
from
Nominal (± V)
0.02 47 ACV 0.2 7.320E-04
0.02 50000 ACV 0.2 1.440E-03
0.2 47 ACV 0.2 Q 1.020E-03
0.1 20 ACV 2 2.200E-03
2 47 ACV 2 2.800E-03
2 10000 ACV 2 2.800E-03
0.1 50000 ACV 2 2.900E-03
2 100000 ACV 2 4.900E-02
20 10000 ACV 20 4.000E-02
20 50000 ACV 20 Q 9.500E-02
200 47 ACV 200 4.100E-01
160 50000 ACV 200 1.230E+00