Calibration
38 Keysight M9181A Calibration and Verification Procedure
3 Perform the calibration process in the SFP: select the Step #, set the source,
enter the Calibrate Value, and click Store Calibration Constant.
Repeat steps 1 through 3 for each gain adjustment point shown in the table
above.
SFP Cal Step Step Name
Calibrator Input
Freq (Hz)
Calibrator Input
Amplitude, Function
7 DCV Gain 0.2 V 0.2 VDC
8 DCV Gain 2 V 2 VDC
9 DCV Gain 20 V 20 VDC
10 DCV Gain 200 V 200 VDC
11 (Skip for M9181A - Applies only to M9182A/83A)
12 ACV Gain 0.2 Vrms 1000 0.2 VAC
13 ACV Gain 2 Vrms 1000 2 VAC
14 ACV Gain 20 Vrms 1000 20 VAC
15 ACV Gain 200 Vrms 1000 200 VAC
16 (Skip for M9181A - Applies only to M9182A/83A)