EasyManua.ls Logo

Keysight M9182A - Verification

Keysight M9182A
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Verification
Keysight M9182A/M9183A Calibration and Verification Procedure 13
Verification
This manual contains procedures for calibrating and verifying the performance of
the Keysight M9182A and M9183A PXI 6½ Digit Multimeters (DMMs).
The following table lists the test equipment recommended for calibration and
performanceverification. You may substitute calibration standards of equivalent
accuracy.
Application Recommended equipment Accuracy requirements
DC Voltage Fluke 5720A <1/5 instrument 24 hour spec
DC Current
Fluke 5720A 10 MW Resistor
[a]
[a] 10 M resistor (25 ppm / °C, >250 V) measured by Keysight 3458A or equivalent
<1/5 instrument 24 hour spec
Resistance Fluke 5720A <1/5 instrument 24 hour spec
AC Voltage Fluke 5720A <1/5 instrument 24 hour spec
AC Current Fluke 5720A <1/5 instrument 24 hour spec
Frequency Fluke 5720A <1/5 instrument 24 hour spec
Capacitance IET SCA–1mF <1/5 instrument 24 hour spec
DCV Source 34410 or equivalent <1/5 instrument 24 hour spec
DCI Source
Fluke 5720A, Keysight 34410A, or
equivalent
<1/5 instrument 24 hour spec

Table of Contents

Other manuals for Keysight M9182A

Related product manuals