Calibration
Keysight M9182A/M9183A Calibration and Verification Procedure 47
Calibration of ACV Flatness
Review “Test Considerations” on page 16 and “Gain Adjustments” on page 44
before beginning this procedure.
Configuration: AC Flatness
1 Configure each function and range shown in the table below.
2 Apply the input signal shown in the “Input” column of the table. Enter the
actual applied rms input voltage (see Adjustments” on page 41).
The next calibration steps perform flatness calibration at 50 kHz on each of the
ranges in the ACV. Turn ACV fast off by selecting slow filtering using the SFP.
Perform the following steps using the “calibration process” described.
3 Perform the calibration process in the SFP: select the Step #, set the source,
enter the Calibrate Value, and click Store Calibration Constant.
Repeat steps 1 through 3 for each gain adjustment point shown in the above
table.
NOTE
Before calibrating the flatness on a specific range, calibrate Gain on that range.
(For example step 17 can be performed only after step 12 is performed).
SFP Cal step Step name
Calibrator input
freq (Hz)
Calibrator input
amplitude, function
17 ACV FComp .2 V 50,000 0.2 VAC
18 ACV FComp 2 V 50,000 2 VAC
19 ACV FComp 20 V 50,000 20 VAC
20 ACV FComp 200/300 V 50,000 160 VAC