Test 4: 2-Wire Ohms Function
26 Keysight M9182A/M9183A Calibration and Verification Procedure
Test 4: 2-Wire Ohms Function
1 Make sure you have read “Test Considerations” on page 16.
2 Set the M9182A/83A DMM to 2-wire ohms function.
3 Connect the calibrator to the input terminals of the DMM, as shown below.
4 On the calibrator, select 2-wire compensation
5 Perform a null measurement with 0 from the calibrator. All 2-wire ohms
measurements are now made relative to this null to remove the lead
resistance.
6 Select each range in the order shown below. Using the calibrator, provide the
resistance value indicated. Compare measurement results to the appropriate
test limits shown in the table. Be certain to allow for appropriate source
settling.