Calibration
Keysight M9182A/M9183A Calibration and Verification Procedure 53
AC Current Gain and Offset Calibration Procedure
Review “Test Considerations” on page 16 and “Gain Adjustments” on page 44
before beginning this procedure.
1 Configure each function and range shown in the table below.
2 Apply the input signal shown in the “Input” column of the table. Enter the
actual applied rms input voltage (see Adjustments” on page 41).
Enter the actual applied input amplitude in Amps rms (see Adjustments” on
page 41).
3 Perform the calibration process in the SFP: select the Step #, set the source,
enter the Calibrate Value, and click Store Calibration Constant.
Repeat steps 1 through 3 for each gain adjustment point shown in the above
table. Store the new calibration constants (see “Storing the Calibration
Constants” on page 41).
NOTE
Before the ACI Offset on a given range can be performed, the ACI gain on that
range must be performed (for example, you must perform step 35 before step
39).
SFP Cal step Step name
Calibrator input
freq (Hz)
Calibrator input
amplitude, function
35 ACI Gain .002 A 1000 0.002 IAC
36 ACI Gain .02 A 1000 0.02 IAC
37 ACI Gain .2 A 1000 0.2 IAC
38 ACI Gain 2 A 1000 2 IAC
39 ACI Offset .002 A 1000 0.0001 IAC
40 ACI Offset .02 A 1000 0.001 IAC
41 ACI Offset .2 A 1000 0.01 IAC
42 ACI Offset 2 A 1000 0.1 IAC