Test 9: Source VDC Performance Verification Tests
Keysight M9182A/M9183A Calibration and Verification Procedure 33
Test 9: Source VDC Performance Verification Tests
1 Make sure you have read “Test Considerations” on page 16.
2 Set the SourceVMeasV function on the M9182A/83A DMM.
3 Connect the 34410A to the input terminals of the DMM as shown below, on the
DCV function.
4 Program each output voltage on the M9182A/83A DMM in the order shown
below. Measure this voltage with the external DMM. Compare these
measurement results to the appropriate test limits shown in the table. Be
certain to allow for appropriate source settling.
34410A
ext DMM
(V)
M9180 series
DMM function
M9180 series
DMM output
voltage (V)
M9180 series
DMM
resolution (V)
Error from
nominal (± V)
10.00000E+00 SourceVMeasV 10 .0001 1.9E-03
1.000000E+00 SourceVMeasV 1 .0001 0.550E-03
-10.00000E+00 SourceVMeasV –10 .0001 1.9E-03