Test 10: Source IDC Performance Verification Tests
34 Keysight M9182A/M9183A Calibration and Verification Procedure
Test 10: Source IDC Performance Verification Tests
1 Make sure you have read “Test Considerations” on page 16.
2 Set the SourceIMeasureV function on the M9182A/83A DMM.
3 For the 10 mA output level, attach the 34410A to the IDC output terminals of
the M9182A/83A DMM as shown below. Put the 34410A in DCI measurement
function. Compare measurement results to the appropriate test limits shown in
the table.
4 For the remaining output levels, connect the 34410A to the input terminals on
the M9182A/83A DMM. Connect the calibrator to the input terminals of the
34410A and set the calibrator to 2-wire . Program the resistance level for
each output level to generate 1 V across the 34410A input leads.
5 Select each output current in the order shown below and provide the indicated
input. Compare measurement results to the appropriate test limits shown in
the table. Be sure to allow for appropriate source settling.