Test 10: Source IDC Performance Verification Tests
Keysight M9182A/M9183A Calibration and Verification Procedure 35
Source
calibrator (Ω)
DMM
function
DMM output
current (A)
Quick check
Error from nominal (± A)
24 hours 90 days 1 year
(N/A) SourceIMeasV 10.00000E-03 182.50E-06 182.50E-06 182.50E-06
1000 SourceIMeasV 1.000000E-03 18.250E-06 18.250E-06 18.250E-06
10,000 SourceIMeasV 100.0000E-06 1.8250E-06 1.8250E-06 1.8250E-06
100,000 SourceIMeasV 10.00000E-06 245.00E-09 245.00E-09 245.00E-09
1,000,000 SourceIMeasV 1.000000E-06 24.500E-09 24.500E-09 24.500E-09