EasyManua.ls Logo

Keysight M9182A - Test 5: Frequency Gain Verification Test

Keysight M9182A
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test 5: Frequency Gain Verification Test
28 Keysight M9182A/M9183A Calibration and Verification Procedure
Test 5: Frequency Gain Verification Test
1 Make sure you have read “Test Considerations” on page 16.
2 On the M9182A/83A DMM, select the Frequency function, Voltage range, and
Frequency range.
3 Connect the calibrator to the input terminals of the DMM, as shown below.
4 Select each range in the order shown below. Using the calibrator, provide the
input voltage and frequency indicated. Compare the frequency measurement
results from the M9182A/83A DMM to the appropriate test limits shown in the
table. Be certain to allow for appropriate source settling.
Calibrator
source input
frequency
(Hz)
Calibrator
source
voltage
input (V)
DMM
voltage
range
(V)
DMM
frequency
range (Hz)
Error from nominal (± Hz)
24 hours 90 days 1 year
10.00000E+00 100.0000E-03 200.0000E-03 2.000000E+00 8.200E-03 8.200E-03 8.200E-03
300.0000E+03 8.000000E+00 20.00000E+00 200.0000E+03 230.0E+00 230.0E+00 230.0E+00

Table of Contents

Other manuals for Keysight M9182A

Related product manuals