Test 4: 2-Wire Ohms Function
Keysight M9182A/M9183A Calibration and Verification Procedure 27
Calibratior
input
resistance (W)
DMM function
DMM
range (W)
Quick
check
Error from nominal (± W)
24 hours 90 days 1 year
19.00000E+00 2W W
20.00000E+00
[a]
2.56E-03 3.51E-03 4.66E-03
190.0000E+00 2W W 200.0000E+00 12.6E-03 24.0E-03 31.7E-03
1.900000E+03 2W W 2.000000E+03 98E-03 193E-03 269E-03
19.00000E+03 2W W 20.00000E+03 Q 971E-03 1.921E+00 2.681E+00
190.0000E+03 2W W 200.0000E+03 15.4E+00 23.0E+00 36.4E+00
1.900000E+06 2W W 2.000000E+06 382E+00 630E+00 820E+00
19.00000E+06 2W W 20.00000E+06 23.20E+03 25.30E+03 38.60E+03
100.0000E+06 2W W
200.0000E+06
[a]
826.0E+03 1.003E+06 1.350E+06
[a] The 20 and 200 M ranges are available only on the M9183A.