Test 3: Ohms Gain Verification Test
Keysight M9182A/M9183A Calibration and Verification Procedure 25
Calibrator
input
resistance (W)
DMM function DMM range (W) Quick check
Error from nominal (± W)
24 hours 90 days 1 year
000.0000E+00 4W W
20.00000E+00
[a]
800E-06 800E-06 1.0E-03
19.00000E+00 4W W
20.00000E+00
[a]
Q 1.56E-03 2.51E-03 3.66E-03
000.0000E+00 4W W 200.0000E+00 4.0E-03 4.0E-03 6.0E-03
190.0000E+00 4W W 200.0000E+00 11.6E-03 23.0E-03 30.7E-03
000.0000E+00 4W W 2.000000E+03 40E-03 40E-03 40E-03
1.900000E+03 4W W 2.000000E+03 Q 97E-03 192E-03 268E-03
000.0000E+00 4W W 20.00000E+03 400E-03 400E-03 400E-03
19.00000E+03 4W W 20.00000E+03 970E-03 1.92E+00 2.68E+00
000.0000E+00 4W W 200.0000E+03 4.0E+00 4.0E+00 4.0E+00
190.0000E+03 4W W 200.0000E+03 15.4E+00 23.0E+00 36.4E+00
000.0000E+00 4W W 2.000000E+06 Q 40E+00 60E+00 60E+00
190.0000E+03 4W W 2.000000E+06 382E+00 630E+00 820E+00
000.0000E+00 4W W 20.00000E+06 400E+00 600E+00 600E+00
19.00000E+06 4W W 20.00000E+06 23.20E+03 25.30E+03 38.60E+03
000.0000E+00 4W W
200.0000E+06
[a]
26.0E+03 30.0E+03 50.0E+03
100.0000E+06 4W W
200.0000E+06
[a]
826.0E+03 1.0300E+06 1.3500E+06
[a] The 20 and 200 M ranges are available only on the M9183A.