Test 1: DC Volts Verification Tests
Keysight M9182A/M9183A Calibration and Verification Procedure 19
A “Q” in the Quick Check column indicates a quick performance verification
test point. All measurements are taken at 32 NPLC
Calibrator
input voltage
(V)
DMM
function
DMM
range (V)
Quick
check
Error from nominal (± V)
24 hours 90 days 1 year
200.0000E-03 DCV 200.0000E-03 Q 7.0E-06 9.6E-06 12E-06
000.0000E+00 DCV 200.0000E-03 Q 1.0E-06 1.6E-06 2E-06
-200.0000E-0
3
DCV 200.0000E-03 7.0E-06 9.6E-06 12E-06
2.000000E+00 DCV 2.000000E+00 44E-06 64E-06 86E-06
000.0000E+00 DCV 2.000000E+00 4E-06 4E-06 6E-06
-2.000000E+0
0
DCV 2.000000E+00 Q 44E-06 64E-06 86E-06
20.00000E+00 DCV 20.00000E+00 920E-06 1.14E-03 1.56E-03
000.0000E+00 DCV 20.00000E+00 120E-06 140E-06 160E-06
-20.00000E+0
0
DCV 20.00000E+00 920E-06 1.14E-03 1.56E-03
200.0000E+00 DCV 200.0000E+00 Q 6.4E-03 8.2E-03 10.6E-03
000.0000E+00 DCV 200.0000E+00 400E-06 400E-06 600E-06
-200.0000E+0
0
DCV 200.0000E+00 6.4E-03 8.2E-03 10.6E-03
300.0000E+00 DCV 300.0000E+00 39.6E-03 69.9E-03 75.9E-03
000.0000E+00 DCV 300.0000E+00 600E-06 900E-06 900E-06
-300.0000E+0
0
DCV 300.0000E+00 39.6E-03 69.9E-03 75.9E-03