Test 2: DC Current Verification Test
Keysight M9182A/M9183A Calibration and Verification Procedure 23
The following examples illustrate how to calculate the proper values in step 8,
above.
If the value of the resistor is 10.02000e6 then the value programmed for the 20
μA range is 20e-6 × 10.02000e6 = 200.4 V.
–For the 2 μA range the value programmed is 2e-6 × 10.02000e6 = 20.04 V, and
for the 200 nA range the value programmed is 200e-9 × 10.02000e6 = 2.004 V.
WARNING
Any voltages greater than 42 V may present a safety hazard. Make certain
that these voltages are not accessible.
Calibrator
source voltage
Calibrator source
IRange (A)
DMM
function
DMM
range (A)
Quick
check
Error from nominal (± A)
24 hours 90 days 1 year
-200.0000E-06 DCI 200.0000E-06 Q 304.0E-09 440.0E-09 600.0E-09
Irange *
Measured 10M
20.00000E-06 DCI 20.00000E-06 10.40E-09 16.60E-09 26.80E-09
Irange *
Measured 10M
-2.000000E-06 DCI 2.000000E-06 1.080E-09 1.660E-09 4.360E-09
Irange *
Measured 10M
200.0000E-09 DCI 200.0000E-09 300.0E-12 366.0E-12 400.0E-12