Test 2: DC Current Verification Test
Keysight M9182A/M9183A Calibration and Verification Procedure 23
The following examples illustrate how to calculate the proper values in step 8,
above.
– If the value of the resistor is 10.02000e6 then the value programmed for the 20
μA range is 20e-6 × 10.02000e6 = 200.4 V.
–For the 2 μA range the value programmed is 2e-6 × 10.02000e6 = 20.04 V, and
for the 200 nA range the value programmed is 200e-9 × 10.02000e6 = 2.004 V.
Any voltages greater than 42 V may present a safety hazard. Make certain
that these voltages are not accessible.
Calibrator
source voltage
Calibrator source
IRange (A)
DMM
function
DMM
range (A)
Quick
check
Error from nominal (± A)
24 hours 90 days 1 year
-200.0000E-06 DCI 200.0000E-06 Q 304.0E-09 440.0E-09 600.0E-09
Irange *
Measured 10M
20.00000E-06 DCI 20.00000E-06 10.40E-09 16.60E-09 26.80E-09
Irange *
Measured 10M
-2.000000E-06 DCI 2.000000E-06 1.080E-09 1.660E-09 4.360E-09
Irange *
Measured 10M
200.0000E-09 DCI 200.0000E-09 300.0E-12 366.0E-12 400.0E-12