Basic Operation
Remote Control
36
Connecting the N9320B to a PC via the LAN Port
N9320B provide you a LAN port to establish an network
conveniently for remote control. Please refer to the following
steps to finish the connection:
1 Switch on the analyzer.
2 Connect the spectrum analyzer to a PC with a LAN cable.
3 Press > {More} > {More} > {IP Admin} > {IP address} to set
IP address for the instrument. For example, set “10.0.0.5” as
the IP address for the instrument.
4 Run Keysight Connection Expert in IO libraries suite.
Right- click on the LAN (TCPIP0) icon, select “Add Instrument’
in the pop- up menu. The “Add LAN Instruments” window
displays for the IP configuration.
5 Select “Add Address”, check “Use IP Address” in the window
and input the IP address as the instrument IP address you
set before.
6 Press “Test Connection” to check the LAN connection. The
figure below indicates that the connection is ready.
Preset/
System