3-4 Keysight N9912A Service Guide
Tests and Adjustments
Before You Begin
3-
Look for metal particles, scratches, and
dents
Connector Cleaning
Do Try compressed air first Do Not Use any abrasives
Use isopropyl alcohol
a
Get liquid into plastic support beads
Clean connector threads
Gaging Connectors
Do Clean and zero the gage before use Do Not Use an out-of-specification
connector
Use the correct gage type
Use correct end of calibration block
Gage all connectors before first use
Making Connections
Do Align connectors carefully Do Not Apply bending force to connection
Make preliminary connection contact
lightly
Over tighten preliminary connection
Turn only the connector nut Twist or screw any connection
Use a torque wrench for final
connection
Tighten past torque wrench “break”
point
a. Cleaning connectors with alcohol shall only be done with the instrument’s power cord removed, and in a well-ventilated
area. Allow all residual alcohol moisture to evaporate, and the fumes to dissipate prior to energizing the instrument.
Table 3-1 Connector Care Quick Reference Guide