4-8 Keysight N9912A Service Guide
Troubleshooting
Measurement Group Troubleshooting
4-
Measurement Group Troubleshooting
If calibration with the standard Type-N cal kit (1250-3607 or N9910X-800)
gives unexpected results:
Make sure the unit is warmed up for at least 30 minutes.
Make sure the correct cal kit definition is being used; erase user data to
force use of factory cal kit definitions (see Chapter 8 for information on the
“Erase User Data Utility”).
If there appears to be LAN communication problems:
Use only 100 Base T LAN connections. Sometimes there are problems with
the slower speed 10 Base T connections.
If “ADC Overload” error messages are displayed during S21 or insertion loss
measurements:
Perform the “ADC Overload Adjustment” under Adjustments (Keysight
N7841A Software Package)” in Chapter 3.