EasyManua.ls Logo

Keysight U1241B - Environment Conditions

Keysight U1241B
83 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight U1241B and U1242B Users and Service Guide 7
Environment Conditions
This instrument is designed for indoor use, in areas with low condensation and to
be used with standard or compatible test probes.
Environment Conditions Requirements
Operating environment Full accuracy at –10 °C to 55 °C
Operating relative humidity
Full accuracy up to 80% RH for temperature up to 30 °C,
decreasing linearly to 50% RH at 55 °C
Storage environment –20 °C to 70 °C
Altitude
0 to 2000 meters per IEC 61010-1 (3rd Edition) CAT III,
1000 V / CAT IV, 600 V
Pollution Degree Pollution Degree II