5 Performance Tests and Calibration
74 Keysight U1241B and U1242B User’s and Service Guide
Valid Adjustment Input Values
Adjustment can be accomplished using the following input values below.
Table 5-3 Valid adjustment input values
Function Range Valid Input Reference Values
1000 mV, 10 V, 100 V, 1000 V 0.9 to 1.1 x Full Scale
1000 mV, 10 V, 100 V, 1000 V 0.9 to 1.1 x Full Scale
1 V 0.9 to 1.1 x Full Scale
Ω 1000 Ω, 10 kΩ, 100 kΩ, 1000 kΩ, 10 MΩ 0.9 to 1.1 x Full Scale
1000 nF, 10 μF, 100 μF, 1000 μF, 10 mF 0.9 to 1.1 x Full Scale
1000 μA, 10000 μA 0.9 to 1.1 x Full Scale
100 mA, 1000 mA 0.9 to 1.1 x Full Scale
10 A 0.9 to 1.1 x Full Scale
T1 0 °C
Ensure that 0 °C with ambient compensation is
provided
DCmV
(T1)
100 mV 0.9 to 1.1 x Full Scale
The minimum AC current output of Fluke 5520A calibrator is 29 μA. Ensure
that at least 50 μA is set at the calibrator source of AC μA.