Performance Tests and Calibration 5
Keysight U1241B and U1242B Users and Service Guide 71
Changing the Multimeter Calibration Security Code from the Front Panel
1 When the multimeter is in the unsecured mode, press button for more
than one second to enter Calibration Security Code setting mode.
2 The factory default calibration security code 1234 will be displayed on primary
display.
3 Press or to step each character in the code. Press or
to change the value of the selected character.
4 Press (Save) button, to store new calibration security code.
5 If the new calibration security code has been successfully stored, the
secondary display will show PASS. If the new code failed to save, the
multimeter will shows error code E07 on the secondary display for
approximately 3 seconds and return to the Calibration Security Code setting
mode.
Unsecuring the Multimeter Without the Security Code
1 Record the last 4 digit serial numbers of the multimeter.
2 Turn the rotary switch to .
3 Press and simultaneously to enter the Calibration Security
Code entry mode. The primary display shows 5555 and the secondary display
shows SEC.
4 Press for more than one second to enter Set Default Security Code
mode. The secondary display shows SEr and primary display shows “5555”.
5 Press or to step each character in the code. Press and
to change the value of the selected character.
6 Set the code, same as the last 4 digit serial number of the multimeter. Press
(Save) to confirm the entry.