Description of the Kingon P2 Oxygen Concentrator
13
Alerts-Continued
Tank Pressure Fail
Pls Contact Provider
Abnormity of the tank pressure ,Concentrator will shut
down after 30 seconds.Pls contact equipment provider
for maintenance
SieveBed Fail
Pls Contact Provider
Sieve bed does not work or become invalid,
Concentrator will shut down after 10 seconds.
Pls contact equipment provider for maintenance
Compressor Fail
Pls Contact Provider
Compressor does not work,
Concentrator will shut down after 10 seconds.
Pls contact equipment provider for maintenance
Valve Check Fail
Pls Contact Provider
Valve failed ,
Concentrator will shut down after 10 seconds.
Pls contact equipment provider for maintenance
Cooling Fan Fail
Pls Contact Provider
Abnormity of the cooling fan,
Concentrator will shut down after 10 seconds.
Pls contact equipment provider for maintenance
Battery Depleted
Replace battery or connect to
adapter
Battery level is <5% when no adapter connected.
Replace a new battery or connect to adapter for charging
System too Cold
Pls Shut down,Move to
warmer place
System temperature is too low
Concentrator will shut down after 10 seconds
Pls move to a warmer place
Battery too Cold
Pls Shut down,Move to
warmer place
Battery temperature is too low
Concentrator will shut down after 10 seconds
Pls move to a warmer place
System too Hot
Shut down,Move to cooler
place
System temperature is too high
Concentrator will shut down after 10 seconds
Pls move to a cooler place
Battery too Hot
Pls Shut down ,Only Use
adapter
Battery temperature is too high
Concentrator will shut down after 10 seconds
Pls move to a cooler place
Gas Supply Fail
Pls Contact Provider
Oxygen was delivered with a lower purity and capacity
compared with the normal situation
Concentrator will shut down after 10 seconds
Sys Startup Fail
Pls Contact Provider
Oxygen purity <87% within 120 seconds after system
start up.
Concentrator will shut down after 10 seconds
Power Supply Fail
Pls Contact Provider
Input Voltage of Power supply can’t ensure the normal
working of concentrator.
Concentrator will shut down after 10 seconds

Other manuals for Kingon P2