Maintenance and Cleaning of Kingon P2
23
.MAINTENANCE AND CLEANING OF KINGON P2
Cleaning the Case
The outside case should be cleaned using a cloth dampened with a solution of mild
detergent and water.
CAUTION
Do not allow liquids into any of the controls, the interior of
the case, or the oxygen tubing connector. If this occurs,
contact your equipment provider for assistance.
WARNING
Cannula Replacement
The nasal cannula is designed for disposable usage. You can buy it from physician
and/or equipment provider and/or cannula manufacturer’s instructions.
CAUTION
Nasal cannula should be rated for 5 liters per minute to
ensure proper patient usage and oxygen delivery.
Filter Cleaning and Replacement
Filters are designed to adequate air flow through the device at the front of the Kingon
P2.
Pre Filter # FI-P200
This particle screen must be cleaned on a weekly basis to ensure
adequate air flow through the device. Clean the pre filters with a mild
liquid detergent and water; be sure that the filter is dried without water
before reuse.
CAUTION
It may be necessary to clean the particle filters more often
in dusty or bad environments/conditions.