System Specifications
27
.SYSTEM SPECIFICATIONS
Concentrator Specification
Dimensions
L/W/H:8.70in.(22.1cm.)/3.35in.(8.5cm.)/6.30in.(16.0cm.)
Weight
4.34 pounds 1.97Kg ( with battery.)
User Interface
2.8 inch large LCD color display screen
Sound Level
49 dB ( A ) ( on setting 2 )
Warm-Up time
2 minutes
Oxygen
Concentration
90% - 3% /+ 6% at all settings
Flow Control
Settings and
Pulse Volumes
Settings
1
2
3
4
5
Breath
Rate
Pulse Volumes(ml)
10
21
42
63
84
100
15
14
28
42
56
66.7
20
10.5
21
31.5
42
50
25
8.4
16.8
25.2
33.6
40
30
7
14
21
28
33.3
35
6
12
18
24
28.6
40
5.3
10.5
15.8
21
25
±15% at STPD*
+/-25% over the rated environmental range
*STPD is 101.3 kPa at an operating temperature of 20 °C,
dry
Breathing
Frequency
10 to 40 BPM
Inspiratory
Trigger
Sensitivity
0.12 cm H2O
Maximum Outlet
Pressure
25 PSI
Use Mode
Continuous Use