Description of the Kingon P2 Oxygen Concentrator
Alerts
Battery Exhausted
Pls contact provider
Battery health condition is less than 50%
Replace Sieve Bed
Pls contact provider
1. molecular sieve is out of date ,remind users to contact
equipment provider to change the sieve bed
2. Chip error of the sieve bed ,remind users to contact
equipment provider for maintenance
Low Input Voltage
Pls Check Adapter
Adapter Input voltage is lower than 17V until the voltage rise to
up 18V
Absence Of Breath
Pls Check Cannula
Concentrator does not detect a breath over 15 seconds
Check if the cannula is connected firmly and not
kinked,correctly weared and breathe using your nose.
Low Oxygen
:
< 87%
Pls Contact Provider
Oxygen purity is lower than 87% for more than 5 minutes
successively.
Pls contact equipment provider for maintenance
Low Battery
Pls Charge Now
Battery level is ≥ 5% and ≤ 20%
pls connect the adapter to charge.
Low Oxygen
:
< 50%
Pls Contact Provider
Oxygen purity is lower than 50% for more than 5 minutes
successively
Concnetrator will shut down after 30 seconds.
Breath Sensor Fail
Pls Contact Provider
Breath sensor does not work ,Concnetrator will shut down
after 30 seconds.Pls contact equipment provider for
maintenance
Oxygen Sensor Fail
Pls Contact Provider
Oxygen sensor does not work,Concentrator will shut down
after 30 seconds.Pls contact equipment provider for
maintenance
Gas Delivery Fail
Pls Contact Provider
Oxygen has been released but not delivered
successfully .Concentrator will shut down after 30 seconds.Pls
contact equipment provider for maintenance
Gas Obstruction
Pls Contact Provider
Output gas tube on the back-end of the tank is blocked
Or the cannula is kinked
Concentrator will shut down after 30 seconds.Pls contact
equipment provider for maintenance