NOTICE!
Danger of damage to the melt water container!
The units in the table below are equipped with a
melt water container on their underside, which can
be damaged by incorrect transport:
– Place unit on pallet and transport it.
– Do not lift unit with forklift or pallet truck unless
it is on a pallet.
BL-520 BL-520-eco
BL-720 BL-720-eco
FROSTER-BL-530
FROSTER-BL-730
Protective equipment:
n
Safety shoes
n
Protective gloves
Decommission the unit as shown in section
Ä
“Final decommis-
sioning of the unit” on page 38.
1. Transport the unit to the new location.
Waiting time before recommissioning:
Once the unit has been transported, set up
the unit and wait one hour before recommis-
sioning, so that the refrigerant can distribute
itself evenly throughout the refrigerant circuit
again.
2.
Recommission the unit as shown in section
Ä
“Putting the
unit back into operation” on page 38.
Inform your specialist dealer of any location
changes!
If the unit is set up at a different location by the
owner or the unit officer, inform the responsible
medical product advisor about the change of loca-
tion.
Transporting the unit to a new loca-
tion
Transport and decommissioning
Unit transport
18.12.2018 Refrigerator/freezer BL series 37