Recommended temperature warning limits
The temperature warning limits must not be the
same as the target temperature.
For blood reserves, set the temperature limits as
follows:
– Upper temperature warning limit: at least 2 °C
higher than the target temperature
– Lower temperature warning limit: at least 2 °C
lower than the target temperature
1. Insert unit key in key switch.
2. Turn unit key to position “P”.
3. Press and hold button [Temperature warning max.].
4. Press button [Max.] repeatedly until the value for the new
upper temperature warning limit is reached.
5. Press button [Min.] repeatedly until the value for the new
lower temperature warning limit is reached.
6. Turn unit key to position “1”.
ð
The Display shows the current interior temperature.
The set temperature warning limits are
saved automatically. The temperature
warning limits are retained after a power
failure and after the unit has been
switched off.
Changing the temperature warning
limits
Commissioning
Programming the unit > Temperature warning limits
18.12.2018 Refrigerator/freezer BL series 45