NOTICE!
For repairs, contact the service department:
The following company is appointed and authorised by
KIRSCH to provide service for the unit:
Ä
‘Service con-
tact’ on page 4
CAUTION!
Danger of death due to interrupted cooling chain
during stock transfer!
The unit is used for commercial storage of pharmaceut-
icals, vaccines and pharmaceutical ingredients that
require cooling in line with the specifications of the
manufacturer. If the cooling chain of the chilled goods is
interrupted by a stock transfer, the prescribed storage
conditions are no longer met.
Do not subject chilled goods to light during the
stock transfer.
Do not place chilled goods near radiators during the
stock transfer.
Make sure that chilled goods are stored in the
replacement unit according to the specifications of
the relevant manufacturer.
NOTICE!
Meaning of “X” for error and status messages
X is not shown on the display.
Instead, the display shows a number that describes
the relevant part.
Tab. 11: Error messages of the unit
Display Buzzer Description Measure
3
Sensor X:
n Error or short circuit in the relevant
sensor.
n The cryostat is running in the emer-
gency program.
n Contact the service department.
3
Sensor X:
n Error or break in the relevant
sensor.
n Cryostat is running in the emer-
gency program.
n Contact the service department.
Status displays and error messages
Error messages
23.05.2023Refrigerator/freezer MED series66