EasyManua.ls Logo

Kirsch MED 340 PRO-ACTIVE - Page 67

Default Icon
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Display Buzzer Description Measure
3
Fan X:
n Speed of the fan in question is too
low.
n The temperature of the chilled
goods may fluctuate.
n Transfer chilled goods to a new
location.
n Contact the service department.
3
Fan X:
n Speed of the relevant fan is too
high.
n The temperature of the chilled
goods may fluctuate.
n Transfer chilled goods to a new
location.
n Contact the service department.
3
Fan:
n Fan does not reach required min-
imum speed after a unit restart.
n The temperature of the chilled
goods may fluctuate.
n Transfer chilled goods to a new
location.
n Contact the service department.
3
Fan:
n Difference between the speeds of
the fans is too great.
n The temperature of the chilled
goods may fluctuate.
n Transfer chilled goods to a new
location.
n Contact the service department.
3
Relay X:
n Defect in the relevant relay.
n The temperature of the chilled
goods may fluctuate.
n Transfer chilled goods to a new
location.
n Contact the service department.
3
Relay X:
n Defect in the relevant relay.
n The temperature of the chilled
goods may fluctuate.
n Transfer chilled goods to a new
location.
n Contact the service department.
3
Synchronisation error:
n Synchronisation error between con-
trol unit and monitoring circuit.
n No secured function of the cooling
controller.
PRO-ACTIVE
n Set key switch to “0”.
n Unplug power plug and switch on
again
Ä
Chapter 7.1 ‘Switching on
the unit’ on page 43.
ULTIMATE
n Shut down
Ä
Chapter 7.2
‘Switching off the unit’ on page 43
n Unplug power plug and switch on
again
Ä
Chapter 7.1 ‘Switching on
the unit’ on page 43.
Status displays and error messages
Error messages
23.05.2023 Refrigerator/freezer MED series 67