INSTALLATION INSTRUCTIONS Installation
10MAR06
II4825-1
Page
42 of 72
Setting the “Push” Option
Important
The “Push” function is determined by the features of the “Broker”.
10 Check the “Verify and Validate” database at the TSC to identify if the “Broker” has the
“Push” function.
11 If the customer wants the “Push” option, select the “Push Connection Active” check box
at each system.
12 Type a value between 10225 - 65535 into the “PACS Push Port Number” field.
Note
The default number is 5401.
The same port number can be used for all systems.
13 Type any name into the “PACS Push AE Title” field except the computer name.
14 Record all the data from the “Push” screen.