INSTALLATION INSTRUCTIONS Installation
10MAR06
II4825-1
Page
59 of 72
Setting the Requirements Fields
1 Obtain from your customer, the required fields to be complete for a patients study.
2 From the main menu, touch:
• [Key Operator Functions]
• [Required Fields]
3 Select the fields obtained in Step 1.
4 Touch:
• [Save]
• [Back]
• [Next]
Configuring the MONITOR
1 Obtain the model information from the back of the MONITOR.
2 From the main menu, touch:
• [Key Operator Functions]
• [System Configuration]
• [Next]
• [Next]
• [Configure Monitor]