EasyManuals Logo

KROHNE BATCHFLUX 3200 C User Manual

KROHNE BATCHFLUX 3200 C
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #3 background imageLoading...
Page #3 background image
SAFETY INSTRUCTIONS 1
3
BATCHFLUX 3200 C
www.krohne.com08/2016 - 4005430201 - MA_Ex_AD_BATCHFLUX 3200 C-
1.1 General notes
These additional instructions apply to explosion-protected versions of electromagnetic
flowmeters in type of protection encapsulation "mb", equipment category II 2 G, EPL Gb. It
completes the installation and operating instructions for the non-explosion protected versions.
The information given in this instruction contains only the data relevant to explosion protection.
The technical details given in the manual for the non-explosion protected versions apply
unchanged unless excluded or superseded by these instructions.
1.2 EU conformity
The manufacturer declares with the EU declaration of conformity, in which the applied standards
are also specified, on his own responsibility conformity with the essential health and safety
requirements of directive 2014/34/EU for equipment used in hazardous areas.
The EU declaration of conformity for the equipment category II 2 G is based on the following EU
type examination certificate:
The applied standards are EN 60079-0:2012 + A11 and EN 60079-18:2015.
The "X" after the certificate number refers to specific conditions for safe use, which are listed in
section "Specific conditions of use".
The EU type examination certificate can be downloaded from the manufacturer's website.
1.3 Approval according to the IECEx scheme
Conformity of the electromagnetic flowmeter for use in hazardous areas with gas was tested in
accordance with the IECEx Certification Scheme for Explosive Atmospheres acc. to IEC 60079-0:
2011 and IEC 60079-18: 2014. The number of the IECEx certificate is:
The "X" after the certificate number refers to specific conditions for safe use, which are listed in
section "Specific conditions of use".
The IECEx certificate can be downloaded from the manufacturer's website.
KIWA 16ATEX0017 X
KIWA 16ATEX0017 XKIWA 16ATEX0017 X
KIWA 16ATEX0017 X
IECEx KIWA 16.0007X
IECEx KIWA 16.0007XIECEx KIWA 16.0007X
IECEx KIWA 16.0007X